6 Laboratory
Quality Assurance System
6.1 Laboratory Construction
In China there are altogether 8 reference laboratories for residue
detection and 31 approved laboratories under AQSIQ, and 22 national
key laboratories, 43 regional central laboratories, 6 routine laboratories
are involved in animal disease detection.
There are 2 reference laboratories for residue detection and 32
approved laboratories under MOA.
6.1.1 reference laboratories
Responsibilities of Reference Laboratories: setting up and coordinating
residue analysis methods and standards; assisting the competent
authorities in establishing and carrying out residue monitoring
programs; participating in international comparative tests and performance
tests; organizing regular comparative tests and performance tests;
collecting and transferring information on international residue
monitoring; offering domestic or international training programs
for analytical staff to improve their analytical skills.
6.1.2 Approved laboratories
Responsibilities of Approved Laboratories: presenting assigned testing
results; strictly carrying out designated standards, methods or
detection limits for residue testing; participating in comparative
tests organized by reference laboratories; proposing to reference
laboratories on improving testing methods.
6.2 Management of Laboratory Quality
All laboratories have been accredited by Laboratory Accreditation
Committee and have adequate quality control documentation in accordance
with ISO/IEC-17025-1997.
The laboratories are finely staffed with employees who have been
trained on residue analysis and are familiar with international
standards and practices. The laboratories are also equipped with
adequate facilities and documentation to complete assigned analysis
projects. The laboratories have good administrative organizations,
own latest CRMs and related information, and possess sufficient
data processing abilities, are able to rapidly submit statistical
data and other information to the competent authorities and at the
same time, notifying other laboratories.
6.3 Laboratory Testing Methods
6.3.1 Routine Methods of analysis
Laboratories use testing methods in accordance with the laws of
inspection and quarantine and the requirements of the national residue
monitoring program. Routine methods of analysis must be verified
in the laboratory and can be used for the purpose of screening and
confirmation.
6.3.1.1 Screening Methods: used for the screening of one kind or
one class of compounds; the sensitivity can meet the requirement
of residue detection. The methods must be suitable for samples of
large numbers for potential positive results; specifically designed
to avoid false negative results.
6.3.1.2 Confirmatory Methods: used for confirmation, which provides
full or complementary information for positive results, specifically
designed to avoid false positive results. The sensitivity can satisfy
the requirements for residue detection.
6.3.2 Interpretation of Results
6.3.2.1 Positive Results: if the presence of a substance that must
not be found in the sample is confirmed, the result is regarded
as "positive". For substance with an established maximum
residue limit, the result of the analysis is considered as ¡®positive¡¯
in case the determined content of the identified compound in the
sample is higher than the MRLs (in any calibration).
6.3.2.2 Negative Results: if a substance that must not be detected
in the sample has proved to be out of existence, the result is regarded
as "negative". For substances with established MRLs, the
result is regarded as ¡®negative¡¯ in case the content of the compound
in the sample is lower than the MRLs.
6.3.3 Choosing of Methods of Analysis
6.3.3.1 Relevant Chinese national standards (GB) or Chinese industrial
standards (SN)
6.3.3.2 Recognized standard methods, for instance, ISO methods
6.3.3.3 Standard Operational Practices (SOP)
authorized by competent authorities or institutions.
6.3.4 Quality Control
6.3.4.1 Analytical methods must be approved or validated thoroughly
before put into use;
6.3.4.2 It has been stipulated that only those who have been examined
and qualified can handle sample analysis, calculation, recording
and checking of results;
6.3.4.3 Use control samples, including certified reference materials
(CRM), or other laboratory reference materials calibrated by the
standard substances recognized internationally. Samples analyzed
each time should be paralleled with quality control samples
6.3.4.4 Recovery, linearity and calibration curve should be checked
by spiked samples (with target component or internal standard).
6.3.4.5 Main testing facilities should meet the requirements of
analytical methods, and be verified and maintained regularly.
6.3.4.6 Data calculation should be presented in detail. The results
must meet statistical quality control requirements.
6.3.5 Participating in Comparative Tests and Performance Tests
Although there are few international tests on residue testing performance,
quite a lot of CIQ laboratories participated in international comparative
tests and performance tests in recent years: 35 laboratories have
been accredited by Japanese Ministry of Welfare and Labor; 9 laboratories
from Zhejiang and Shanghai took part in tea residue detection performance
tests organized by TPA and have been accredited by European Tea
Committee; Laboratories from Liaoning, Zhejiang and Shanghai participated
in performance tests organized by APLAC; commodity inspection research
institutes in Qinhuangdao and Dalian participated in AOAC testing
method co-experiment (on items of Clopitol, pyrethroids, etc.) In
the meantime, State Laboratory Accreditation Committee frequently
organizes nationwide laboratory performance tests, blind sample
tests and comparative tests.
|